Start typing and press Enter to search

This website does not support Internet Explorer. For a correct visualization we recommend to use Microsoft Edge or Google Chrome.

Optical Metrology

Stent optical inspection system calibration and performance

Optical Metrology
研究開発エンジニアリングマネージャー、光工学博士、フォトニクス理学修士 at Sensofar Metrology | Other articles

2010年よりSensofarで共焦点法、干渉法、焦点移動法の開発に携わる。2018年より研究開発エンジニアリングマネージャーを務める。光学機械システム設計と画像処理に焦点を当てている。\n光工学におけるその総合的な研究活動は、Sensofarの研究開発グループをイノベーションと高い技術レベルで常に最先端に立つ卓越した位置付けへと導いている。

Stent optical inspection system calibration and performance full article
C. Bermudez,1,2 F. Laguarta,1 C. Cadevall,1,2 A. Matilla,2 S. Ibañez,3 and R. Artigas1,2
1Center for Sensors, Instruments and Systems Development (CD6), Universitat Politècnica de Catalunya (UPC) Rambla Sant Nebridi, 10, E-08222 Terrassa, Spain.
2Sensofar-Tech, S.L., (Spain)
3Sensofar-Medical, S.L., (Spain)
Applied Optics Vol. 56, Issue 9,pp. D134-D141 (2017)

Abstract

Implantable medical devices, such as stents, have to be inspected 100% so no defective ones are implanted into a human body. In this paper, a novel optical stent inspection system is presented. By the combination of a high numerical aperture microscope, a triple illumination system, a rotational stage, and a CMOS camera, unrolled sections of the outer and inner surfaces of the stent are obtained with high resolution at high speed with a line-scan approach. In this paper, a comparison between the conventional microscope image formation and this new approach is shown. A calibration process and the investigation of the error sources that lead to inaccuracies of the critical dimension measurements are presented.

References

Carlos Bermudez, Ferran Laguarta, Cristina CadevallAitor Matilla, Sergi Ibañez, and Roger Artigas. Optical stent inspection of surface texture and coating thickness, Proc. SPIE 10110, Photonic Instrumentation Engineering IV, 1011007 (20 February 2017)

Carlos Bermudez, Ferran Laguarta, Cristina Cadevall, Aitor Matilla, Sergi Ibañez and Roger Artigas. Novel Stent Optical Inspection System, 2016, Applied Industrial Optics: Spectroscopy, Imaging and Metrology, Paper# AITh2B.3

Contact Us

We're not around right now. But you can send us an email and we'll get back to you, asap.

Start typing and press Enter to search

Confocal fusion3D confocal measurements of turning parts