Measurement modes
Imaging

Screenshot

Unrolled FOV

Unrolled section
3D Modes

Surface topography

Surface roughness

Coating thickness
Inspection capabilities

Outer surface

Inner surface

Sidewalls
Measurable stent
Material | Metal (Stainless Steel, CoCr) and Nitinol |
OD | 1 – 15 mm |
Length | Up to 100mm (XL stage option 250mm) |
Light sources

530nm

550nm
Dimensions


Sensorhead & Controller
Objectives performance
MAG | 2.5X | 5X | 10X | 20X |
Numerical Aperture | 0.08 | 0.15 | 0.20 | 0.45 |
Working Distance (mm) | 8.80 | 23.50 | 37.00 | 4.50 |
Horizontal FOV (mm) | 7.20 | 3.60 | 1.80 | 0.90 |
Spatial sampling (µm)[1] | 3.52 | 1.76 | 0.88 | 0.44 |
Optical resolution (µm)[2] | 1.95 | 0.97 | 0.73 | 0.32 |
Acquisition time (mm2/s) | 20 | 5 | 1.25 | 0.31 |
1 Pixel size on the imaged surface. 2 L&S: Line and Space, half of the diffraction limit according to the Rayleigh criterion. Values for blue LED.
MAG | 10XDI | 20XDI | 50XDI |
Numerical Aperture | 0.30 | 0.40 | 0.55 |
Working Distance (mm) | 7.40 | 4.70 | 3.40 |
Horizontal FOV (mm) | 1.80 | 0.90 | 0.36 |
Spatial sampling (µm)[1] | 0.88 | 0.44 | 0.18 |
Optical resolution (µm)[2] | 0.48 | 0.36 | 0.26 |
Vertical resolution (nm)[3] | 1 | 1 | 1 |
1 Pixel size on the imaged surface. 2 L&S: Line and Space, half of the diffraction limit according to the Rayleigh criterion. Values for blue LED.
3 Vibration isolation table is required.
Computer & Operating System

Computer | HP Z220 SFF, HP display 2560×1440 pixels |
Operating System | Windows 10, 64 bits |
Data management

Depending on the intended use, the Q six can be operated in three different environments: Development, Quality Assurance (QA) and Production. Each environment has its own databases.
Data is organized in three different categories:
1 System Databases containing User data and System configuration files
2 Stent Inspection Databases containing Stent Model data, Defect libraries and Acquisition, Analysis and Report recipes
3 Inspection Results

Sensofar Medical can provide specialized support to QA/RA departments in qualification and validation processes. A complete validation package for the Q six is available as an option.