Round Robin test on bio-imaging transfer standard for 3D optical profilers

Optische Metrologie
R&D Engineering Manager, Doktor in Optik, MSc in Photonik at Sensofar Metrology | Other articles

Carlos beschäftigt sich bei Sensofar seit 2010 mit der Entwicklung von Konfokal-, Interferometrie- und Fokusvariationstechnologien, seit 2018 hat er die Position des R&D Engineering Manager inne. Seine Interessen sind optomechanisches Systemdesign und Bildverarbeitung.
Fundierte Forschungsarbeit im Bereich der optischen Technik verleiht der Sensofar F&E-Gruppe eine herausragende Position, um in Bezug auf Innovation und höchstes technologisches Niveau immer auf dem neuesten Stand zu bleiben.

Round Robin test on bio-imaging transfer standard for 3D optical profilers, full article
A. Nolvi,1,2 T. Viitala,1 A. García Pérez,1 N. Sandler,3 E. Hæggström,1 C. Bermudez,4 R. Artigas,4 I. Kassamakov1
1Univ. of Helsinki (Finland)
2Åbo Akademi (Finland)
3Åbo Akademi Univ. (Finland)
4Sensofar Tech SL, (Spain)

Proceedings Volume 10110, Photonic Instrumentation Engineering IV; 101100M (2017)
Event: SPIE OPTO, 2017, San Francisco, California, United States

Abstract

A stair case height Bio-Transfer-Standard (BTS), developed and produced at the University of Helsinki (UH), was measured in two laboratories. The Round Robin test on bio-imaging, aims to determine whether BTS works with different optical profilers in different laboratories. First the artefact was measured at UH using a custom-built Scanning White Light Interferometer. Then BTS was measured at Sensofar-Tech, S.L. using an S neox type interferometer working either in Phase Shifting Interferometry mode or in Imaging Confocal Microscopy mode. To remove the influence of system calibration, a method featuring sample shifting and measurement subtraction was used. The BTS features eight lipid bilayer steps that each are 4.6 ± 0.1 nm tall on average.

All 30 measurements done by four different operators at the two laboratories agree to within 0.1 nm which agrees with theoretical estimates and with measurements done using a surface plasmon resonance technique. The Round Robin results show the applicability of the newly developed bio-imaging transfer standard for calibrating 3D optical profilers.