Metrological characterization of different methods for recovering the optically sectioned image by means of structured light full article
P.Martínez1, C.Bermudez1, G.Carles1, C.Cadevall1, A. Matilla1, J.Marine1, R.Artigas1
1Sensofar-Tech, S.L. (Spain)
Proceedings Volume 11782, Optical Measurement Systems for Industrial Inspection XII; 117820Q (2021) https://doi.org/10.1117/12.2592371
Event: SPIE Optical Metrology, 2021, Online Only
Abstract
Imaging confocal microscopy (ICM) and focus variation (FV) are two of the most used technologies for 3D surface metrology. Both methods rely on the depth of focus of the microscope objective, which depends on its numerical aperture and wavelength of the light source to compute an optical section. In this paper we study how several methods of structured illumination microscopy affect the metrological characteristics of an areal optical profiler. We study the effect of the projection of different structured patterns, the sectioning algorithms, and the use of high and low frequency components onto the optically sectioned image. We characterized their performance in terms of system noise, instrument transfer function and metrological characteristics such as roughness parameters and step height values.
Other publications
3D shape and texture measurements on the micro and nano-scales using a new dual-technology optical sensor head
We introduce a new dual-technology (confocal & interferometer) illumination hardware setup, in this paper.
Testing and applicability of the UPC-ZEBRA interferometer as a phasing system in segmented-mirror telescopes
We present the most representative results obtained during the testing stage of an instrument called UPC-ZEBRA that uses a novel interferometric [...]