Metrological characterization of different methods for recovering the optically sectioned image by means of structured light full article
P.Martínez1, C.Bermudez1, G.Carles1, C.Cadevall1, A. Matilla1, J.Marine1, R.Artigas1
1Sensofar-Tech, S.L. (Spain)
Proceedings Volume 11782, Optical Measurement Systems for Industrial Inspection XII; 117820Q (2021) https://doi.org/10.1117/12.2592371
Event: SPIE Optical Metrology, 2021, Online Only
Abstract
Imaging confocal microscopy (ICM) and focus variation (FV) are two of the most used technologies for 3D surface metrology. Both methods rely on the depth of focus of the microscope objective, which depends on its numerical aperture and wavelength of the light source to compute an optical section. In this paper we study how several methods of structured illumination microscopy affect the metrological characteristics of an areal optical profiler. We study the effect of the projection of different structured patterns, the sectioning algorithms, and the use of high and low frequency components onto the optically sectioned image. We characterized their performance in terms of system noise, instrument transfer function and metrological characteristics such as roughness parameters and step height values.
Other publications
Residual flatness error correction in three-dimensional imaging confocal microscopes
In this paper we show that the residual flatness error on a reference sphere measured with a 10X can make the measurement of the radius to [...]
Active Illumination Focus Variation
In this paper, we show the development of an Active illumination Focus Variation on an existing